Generic placeholder image

Current Physical Chemistry

Editor-in-Chief

ISSN (Print): 1877-9468
ISSN (Online): 1877-9476

Research Article

Residual Moisture Determination in Lyophilized Drug Product Samples Using FT-NIR: A Comparison of Data Processing Methods

Author(s): R. Burtea, T. Upton, G.M. Jensen, S. Kim and K.J. Jalkanen*

Volume 14, Issue 1, 2024

Published on: 22 December, 2023

Page: [67 - 83] Pages: 17

DOI: 10.2174/0118779468275339231206062433

Price: $65

Abstract

Introduction: Remdesivir for injection is a Gilead Sciences manufactured drug product approved by the FDA to treat COVID-19. One of the critical attributes of this drug product, which is lyophilized, is its water moisture content. The current method used to determine the water content of remdesivir is the Karl Fischer titration method. The Karl Fischer method is considered the gold standard for determining the residual water moisture content of lyophilized drug products in the pharmaceutical industry. However, it is timeconsuming, can at times be sensitive to difficult-to-control techniques, and most importantly is destructive. The purpose of this study was to create a non-destructive residual water moisture determination method using FT-NIR.

Methods: Three FT-NIR methods were proposed as a non-destructive alternative to Karl Fischer method: partial least squares regression, Beer’s law and area under the curve, and Beer’s law and peak height. Results were statistically validated using RMSEC, RMSEP, and R2, and the percent differences of predicted and accepted residual moisture values were compared with external validation vials.

Results: Results showed that the partial least squares regression method produced a 0.999 correlation coefficient and a 95% performance index score, highlighting the accuracy between the actual Karl Fischer moisture values and the predicted FT-NIR moisture values. The Beer’s law methods produced acceptable correlation coefficients of 0.996 and 0.982 for the peak area and height, respectively, but comparisons between the actual and predicted values were not as closely matched as the partial least squares method.

Conclusion: Our results support the use of a non-destructive partial least squares FT-NIRbased moisture contest method, which is a useful alternative to the destructive Karl Fischer method for determining residual water moisture of lyophilized drug products. Only an FTNIR instrument and chemometric statistical software along with the appropriately trained/educated analysts, and routine Karl Fischer instrumentation to measure/determine moisture content values of either water spiked or lyo-thieved and normal lyophilized samples are required.

« Previous
Graphical Abstract

[1]
Tavčar E, Turk E, Kreft S. Simple modification of karl-Fischer titration method for determination of water content in colored samples. J Anal Methods Chem 2012; 2012: 1-6.
[http://dx.doi.org/10.1155/2012/379724] [PMID: 22567558]
[2]
Derksen MWJ, van de Oetelaar PJM, Maris FA. The use of near-infrared spectroscopy in the efficient prediction of a specification for the residual moisture content of a freeze-dried product. J Pharm Biomed Anal 1998; 17(3): 473-80.
[http://dx.doi.org/10.1016/S0731-7085(97)00216-1] [PMID: 9656158]
[3]
Yip WL, Gausemel I, Sande SA, Dyrstad K. Strategies for multivariate modeling of moisture content in freeze-dried mannitol-containing products by near-infrared spectroscopy. J Pharm Biomed Anal 2012; 70: 202-11.
[http://dx.doi.org/10.1016/j.jpba.2012.06.043] [PMID: 22824636]
[4]
Bruttel P, Schlink R. Water Determination by Karl Fischer Titration. Metrohm Ion Analysis 2003; pp. 4-62.
[5]
[6]
Agency for Toxic Substances and Disease Registry. Public Health Statement Pyridine CAS#: 110-86-1. Division of Toxicology 1992.
[7]
Thermo Fisher Scientific. Determination of moisture content in freeze-dried materials by FT-nir Available from: https://assets.thermofisher.com/TFS-Assets/CAD/Application-Notes/D14186.pdf
[8]
Brülls M, Folestad S, Sparén A, Rasmuson A, Salomonsson J. Applying spectral peak area analysis in near-infrared spectroscopy moisture assays. J Pharm Biomed Anal 2007; 44(1): 127-36.
[http://dx.doi.org/10.1016/j.jpba.2007.02.013] [PMID: 17391888]
[9]
Mantanus J, Ziémons E, Lebrun P, et al. Moisture content determination of pharmaceutical pellets by near infrared spectroscopy: Method development and validation. Anal Chim Acta 2009; 642(1-2): 186-92.
[http://dx.doi.org/10.1016/j.aca.2008.12.031] [PMID: 19427475]
[10]
Rinnan Å, Berg F, Engelsen SB. Review of the most common pre-processing techniques for near-infrared spectra. Trends Analyt Chem 2009; 28(10): 1201-22.
[http://dx.doi.org/10.1016/j.trac.2009.07.007]
[11]
Zheng Y, Lai X, Bruun SW, et al. Determination of moisture content of lyophilized allergen vaccines by NIR spectroscopy. J Pharm Biomed Anal 2008; 46(3): 592-6.
[http://dx.doi.org/10.1016/j.jpba.2007.11.011] [PMID: 18162356]
[12]
Abdi H. Partial least squares regression and projection on latent structure regression (PLS Regression). Wiley Interdiscip Rev Comput Stat 2010; 2(1): 97-106.
[http://dx.doi.org/10.1002/wics.51]
[13]
Clavaud M, Lema-Martinez C, Roggo Y, et al. Nearinfrared spectroscopy to determine residual moisture in freeze-dried products: model generation by statistical design of experiments. J Pharm Sci 2020; 109(1): 719-29.
[http://dx.doi.org/10.1016/j.xphs.2019.08.028] [PMID: 31499067]
[14]
Thekkudan DF, Rutan SC, Carr PW. A study of the precision and accuracy of peak quantification in comprehensive two-dimensional liquid chromatography in time. J Chromatogr A 2010; 1217(26): 4313-27.
[http://dx.doi.org/10.1016/j.chroma.2010.04.039] [PMID: 20462587]
[15]
Jensen G, Kim Y, Lee KY, Pham H, Upton T, Yang S. S. Phospholipid formulations of 1’-cyano substituted carba-nucleoside analogs. U.S. Patent 17824288, 2023.
[16]
Karunathilaka SR, Fardin-Kia AR, Roberts D, Mossoba MM. Determination of moisture in olive oil: Rapid FT-NIR spectroscopic procedure based on the karl-fischer reference method. J Oleo Sci 2020; 69(11): 1373-80.
[http://dx.doi.org/10.5650/jos.ess20078] [PMID: 33055436]
[17]
Krauklis A, Gagani A, Echtermeyer A. Near-infrared spectroscopic method for monitoring water content in epoxy resins and fiber-reinforced composites. Materials 2018; 11(4): 586.
[http://dx.doi.org/10.3390/ma11040586] [PMID: 29641451]
[18]
Avila CR, Ferré J, de Oliveira RR, et al. Process monitoring of moisture content and mass transfer rate in a fluidised bed with a low cost inline MEMS NIR sensor. Pharm Res 2020; 37(5): 84.
[http://dx.doi.org/10.1007/s11095-020-02787-y] [PMID: 32318827]
[19]
Peng D, Liu Y, Yang J, Bi Y, Chen J. Nondestructive detection of moisture content in walnut kernel by near-infrared diffuse reflectance spectroscopy. J Spectrosc 2021; 2021: 1-9.
[http://dx.doi.org/10.1155/2021/9986940]
[20]
Fardin-Kia AR, Karunathilaka SR, Yakes BJ, et al. A rapid, univariate FT-NIR procedure to determine moisture concentration in olive oil. J Oleo Sci 2019; 68(11): 1105-12.
[http://dx.doi.org/10.5650/jos.ess19118] [PMID: 31695015]
[21]
Windig W, Shaver J, Bro R. Loopy MSC: A simple way to improve multiplicative scatter correction. Appl Spectrosc 2008; 62(10): 1153-9.
[http://dx.doi.org/10.1366/000370208786049097] [PMID: 18926026]
[22]
Zaid A, Abu-Khalaf N, Mudalal S, Petracci M. Differentiation between normal and white striped turkey breasts by visible/near infrared spectroscopy and multivariate data analysis. Food Sci Animal Resour 2019; 40.
[http://dx.doi.org/10.5851/kosfa.2019.e88]
[23]
Maleki MR, Mouazen AM, Ramon H, De Baerdemaeker J. Multiplicative scatter correction during on-line measurement with near infrared spectroscopy. Biosyst Eng 2007; 96(3): 427-33.
[http://dx.doi.org/10.1016/j.biosystemseng.2006.11.014]
[24]
TQ Analyst Software User Guide. Thermo Scientific
[25]
Shao X, Cui X, Wang M, Cai W. High order derivative to investigate the complexity of the near infrared spectra of aqueous solutions. Spectrochim Acta A Mol Biomol Spectrosc 2019; 213: 83-9.
[http://dx.doi.org/10.1016/j.saa.2019.01.059] [PMID: 30684883]
[26]
Davies AMC. Back to basics: Spectral pre-treatments - Derivatives. Spectroscopy Europe 2007; 19: 32-3.
[27]
Miyazawa M, Sonoyama M. Second derivative near infrared studies on the structural characterisation of proteins. J Near Infrared Spectrosc 1998; 6(A): A253-7.
[http://dx.doi.org/10.1255/jnirs.204]
[28]
Dehghani H, Leblond F, Pogue BW, Chauchard F. Application of spectral derivative data in visible and near-infrared spectroscopy. Phys Med Biol 2010; 55(12): 3381-99.
[http://dx.doi.org/10.1088/0031-9155/55/12/008] [PMID: 20505221]
[29]
Alcantara IM, Naranjo J, Lang Y. Model selection using PRESS statistic. Comput Stat 2023; 38(1): 285-98.
[http://dx.doi.org/10.1007/s00180-022-01228-1]
[30]
Glossary for RESULT software and Antaris systems. Thermo Scientific 2023.
[31]
Chapter 15 Quantitative Analysis By Gas Chromatography Measurement of Peak Area and Derivation of Sample Composition. J Chromat Library 1988; 42: 629-59.
[http://dx.doi.org/10.1016/S0301-4770(08)70087-3]

Rights & Permissions Print Cite
© 2024 Bentham Science Publishers | Privacy Policy