Generic placeholder image

Medicinal Chemistry

Editor-in-Chief

ISSN (Print): 1573-4064
ISSN (Online): 1875-6638

Description of High Purity and High Specific Activity of [11C]Choline Synthesis Using TRACERlab FXc Module, and Detailed Report of Quality Controls

Author(s): Giorgio Biasiotto, Francesco Bertagna, Ugo Biasiotto, Carlo Rodella, Giovanni Bosio, Luigi Caimi, Giovanni Bettinsoli and Raffaele Giubbini

Volume 8, Issue 6, 2012

Page: [1182 - 1189] Pages: 8

DOI: 10.2174/1573406411208061182

Price: $65

Abstract

We proposed a method of synthesis to produce [11C]Choline using TRACERlab FXc module that utilized gas phase iodination. The product had radiochemical purity of 99.79 + 0.14 % and specific activity of 45.7 + 7.59 GBq/mol. [11C]Choline did not have at the moment a specific monograph in European Pharmacopeia therefore we used, when possible, as quality controls reference the monograph of [18F]FDG and we proposed suitable methods to verify radiochemical purity and to quantify residual DMAE and choline amounts.

Keywords: [11C]CH3I, [11C]Choline, gas phase iodination, prostate cancer, PET, quality controls, radiopharmaceutical, TRACERlab FXc


Rights & Permissions Print Cite
© 2024 Bentham Science Publishers | Privacy Policy