Generic placeholder image

Current Diabetes Reviews

Editor-in-Chief

ISSN (Print): 1573-3998
ISSN (Online): 1875-6417

Continuous Glucose Monitoring: Physiological and Technological Challenges

Author(s): Gerard Reach and Carine Choleau

Volume 4, Issue 3, 2008

Page: [175 - 180] Pages: 6

DOI: 10.2174/157339908785294488

Price: $65

Abstract

Over the past decade, several continuous glucose monitoring systems have been developed, representing remarkable technological achievements. Most of the systems monitor glucose invasively in the subcutaneous tissue. It is important to realize that there are discrepancies between blood and interstitial glucose concentration, which (1) may impact the quality of the system calibration and thereby the accuracy of the data, (2) may jeopardize the specificity and the sensitivity of hypoglycaemic alarms based on these systems and (3) must be considered in the design of closed-loop insulin delivery systems. The aim of this review is to make the point that the challenge of developing a continuous glucose monitoring system is not only technological, but must also take into account the physiology of glucose in alternate sites where it is sensed.

Keywords: Continuous glucose monitoring, Interstitial glucose, Push-pull phenomenon


Rights & Permissions Print Cite
© 2024 Bentham Science Publishers | Privacy Policy