Abstract
Background: Electrical impedance tomography (EIT) is an imaging modality which is getting more attention these days. The main advantages are real time images, small cost, small size and non-ionizing operation. On the other hand, the biggest drawbacks are blurred images obtained and time consuming implementation.
Objective: This paper introduces challenges and possible remedies when the electrical impedance tomography (EIT) is applied to a real patient. EIT is a promising method for patient real-time monitoring of thorax.
Method: The application of electrodes is time consuming and very sensitive to proper positioning. We propose a method to improve such drawbacks. It enables a pseudo-absolute monitoring, which is superior to classic differential monitoring through the use of VERA device (variable electrical ring analyzer).
Results: This superiority has several aspects. The most important is that physician can see whole lungs and possible other features whereas traditional method is capable of monitoring only differences between the inhalation and exhalation of air. Moreover it is possible to see also anomalities which do not change between the course of breathing - possible cancer detection. Other advantages are faster implementation on patient body, which is currently a huge drawback of traditional EIT electrodes positioning and inherent more reliable obtained images since the proposed construction is a well defined object with certain static dimensions. Another advantage comes from the fact that most of the measurement anomalities are near the measuring electrodes and since the electrodes are not in direct contact with a patient body, the anomalities do not appear in the image of human cross section.
Conclusion: The proposed method is a viable imaging modality which can be used for imaging predominantly lungs providing better insight into what is happening in real time. We hope that such device can contribute to widespread of EIT, which can be more pleasant to patients than traditional computed tomography or magnetic resonance and, at the same time, provide medicals with all necessary information.
Keywords: EIDORS, electrical impedance tomography, health care, Labview, Netgen, real-time monitoring.
Graphical Abstract
Current Respiratory Medicine Reviews
Title:Improved Approach to Electrical Impedance Tomography through Pseudo-Absolute Imaging Method
Volume: 13 Issue: 4
Author(s): David Krcmarik*, Michal Petru and Ivan Masin
Affiliation:
- Institute for Nanomaterials, Advanced Technologies and Innovation, Technical University of Liberec, Studentska 2, 461 17, Liberec 1,Czech Republic
Keywords: EIDORS, electrical impedance tomography, health care, Labview, Netgen, real-time monitoring.
Abstract: Background: Electrical impedance tomography (EIT) is an imaging modality which is getting more attention these days. The main advantages are real time images, small cost, small size and non-ionizing operation. On the other hand, the biggest drawbacks are blurred images obtained and time consuming implementation.
Objective: This paper introduces challenges and possible remedies when the electrical impedance tomography (EIT) is applied to a real patient. EIT is a promising method for patient real-time monitoring of thorax.
Method: The application of electrodes is time consuming and very sensitive to proper positioning. We propose a method to improve such drawbacks. It enables a pseudo-absolute monitoring, which is superior to classic differential monitoring through the use of VERA device (variable electrical ring analyzer).
Results: This superiority has several aspects. The most important is that physician can see whole lungs and possible other features whereas traditional method is capable of monitoring only differences between the inhalation and exhalation of air. Moreover it is possible to see also anomalities which do not change between the course of breathing - possible cancer detection. Other advantages are faster implementation on patient body, which is currently a huge drawback of traditional EIT electrodes positioning and inherent more reliable obtained images since the proposed construction is a well defined object with certain static dimensions. Another advantage comes from the fact that most of the measurement anomalities are near the measuring electrodes and since the electrodes are not in direct contact with a patient body, the anomalities do not appear in the image of human cross section.
Conclusion: The proposed method is a viable imaging modality which can be used for imaging predominantly lungs providing better insight into what is happening in real time. We hope that such device can contribute to widespread of EIT, which can be more pleasant to patients than traditional computed tomography or magnetic resonance and, at the same time, provide medicals with all necessary information.
Export Options
About this article
Cite this article as:
Krcmarik David*, Petru Michal and Masin Ivan , Improved Approach to Electrical Impedance Tomography through Pseudo-Absolute Imaging Method, Current Respiratory Medicine Reviews 2017; 13 (4) . https://dx.doi.org/10.2174/1573398X14666180215160959
DOI https://dx.doi.org/10.2174/1573398X14666180215160959 |
Print ISSN 1573-398X |
Publisher Name Bentham Science Publisher |
Online ISSN 1875-6387 |

- Author Guidelines
- Bentham Author Support Services (BASS)
- Graphical Abstracts
- Fabricating and Stating False Information
- Research Misconduct
- Post Publication Discussions and Corrections
- Publishing Ethics and Rectitude
- Increase Visibility of Your Article
- Archiving Policies
- Peer Review Workflow
- Order Your Article Before Print
- Promote Your Article
- Manuscript Transfer Facility
- Editorial Policies
- Allegations from Whistleblowers