Generic placeholder image

Current Pharmaceutical Analysis

Editor-in-Chief

ISSN (Print): 1573-4129
ISSN (Online): 1875-676X

Vancomycin Microbial Assay Using Kinetic-Reading Microplate System

Author(s): Tulia de Souza Botelho, Felipe Rebello Lourenco and Terezinha de Jesus. Andreoli Pinto

Volume 9, Issue 2, 2013

Page: [172 - 176] Pages: 5

DOI: 10.2174/1573412911309020007

Price: $65

Abstract

The aim of this work was to develop a microbiological assay of vancomycin employing microplates and kinetic reading mode, and to validate the developed method by assessing of specificity and selectivity, linearity, linear range, accuracy and precision parameters. Established conditions comprised a vancomycin standard-curve in concentrations from 0.6 to 1.0 μg mL-1, and soybean-casein broth inoculated with 5% of Bacillus subtilis (ATCC 6633) suspension. Satisfactory results were obtained after a 5 hour incubation. The developed method showed appropriate selectivity and linearity in the range from 0.6 to 1.0 μg mL-1, respectively, satisfactory accuracy (recovery = 97.0%) and precision (RSD = 7.8%).

Keywords: Agar diffusion, Antibiotic assay, Glycopeptide antibiotic, Kinetic reading, Method development, Method validation, Microbial assay, Microplate assay, Turbidimetric assay, Vancomycin


Rights & Permissions Print Cite